ForTest -T2710, Digital Portable Calibrator
- Black anodized aluminum case
- Rechargeable lithium io battery
- LCD display with wide viewing
- Three usage button
- Automatic Auto zero and from keyboard
- Working temperature 15+40 °C
- Automatic sleep mode
- Pressure full scale: (mbar) 100
- Flow Scale:30 or 200 cc/min
The T2710 is the digital portable calibrator for micro flow of ForTest® T-Series.
Enclosed in a container of anodized aluminium with shatterproof plexiglass front, we find a completely new circuit board and a mass flow meter sensor, for direct measurement of leaks.
Coupled to the micrometer nozzle supplied, they are designed to simulate the desired leakage, facilitating the procedure for programming and verification of leak test instruments.
The T2710 Standard version has got a lithium ion battery rechargeable that gives it an incredible autonomy and a USB port for PC connection in order to download and manage the taken measures.
The T2710 uses a mass flow sensor and can detect flow rates up to 200 cc/min with precision and reliability. This portable instrument, of compact dimensions and simple use is essential to check the calibration and the repeatability of measure instruments and of nozzle sample.
Type | Flow Calibrator |
Resolution | 0.01 sccm/min |
Data Storage | USB |
Accuracy | 0.5% |
Testing Gas
Precision micrometer nozzle | |
Micrometer nozzle | |
Serial communication using USB cable (ForTest protocol) | |
High resolution measure - (STANDARD) |
T2710 digital calibrator Datasheet |
- Weight 28g
- Accuracy ±5% + 2 cc/min
- Max working pressure 10 bar (over 10 bar under request)
- Flow values starting from 200 cc/min
- Made of stainless steel
- SIT certificated calibration
- Stäubli© male connector
- Usable for pressure and vacuum applications
Accuracy | 0.5% |
- Aluminium Probe, with a 30° arm
- Warning led and buzzer
- Non-explosive, 5% hydrogen mix
- USB slave, RS232, RS485, Can bus and TTY
Type | Mass Flow |
# of Channels | 1 |
Resolution | 1 Pa (0.00015 psi) |
Data Storage | USB |
Interface | RS-232, RS-485 |
Accuracy | 0.5% |
- Weight 28g
- Accuracy ±5% + 0,03 cc/min
- Max working pressure 10 bar
- Leak scales from 10 cc/h to 200 cc/min
- Made of stainless steel
- SIT certificated calibration
- Staubli male connector
- Usable for pressure and vacuum applications
Accuracy | 0.5% |
- Multiple Programs for various applications
- HMI touchscreen controller
- Multiple Languages, English, French, Spanish, Italian, …
- Sequence – Fill to Test Pressure, Stabilize, Measure Absolute Pressure Decay, calibrated leak rate
- Data Collection by Serial Number, using a PC or PLC (optional Field buses, Ethernet IP, Modbus)
Type | Mass Flow |
# of Channels | 1 |
Resolution | 0.01 sccm/min |
Display | 7" Colour Display |
Data Storage | USB |
Interface | RS-232, RS-485, Profinet, EtherCAT, Ethernet TCP/IP |
- Fully automatic test flow adjustment management (8,000 Lt/min)
- Multiple Programs for different parts
- HMI touchscreen controller
- Multiple Languages, English, French, Spanish, Italian, …
- Sequence – Fill to Test Pressure, Stabilize, Measure Absolute Pressure Decay, calibrated leak rate
- Data Collection by Serial Number, using a PC or PLC (optional Field buses, Ethernet IP, Modbus)
Type | Mass Flow |
# of Channels | 1 |
Resolution | 0.01 sccm/min |
Data Storage | USB |
Interface | RS-232, RS-485, Profinet, EthernetIP, EtherCAT, Ethernet TCP/IP |
Accuracy | 0.5% |
- Electronically Controlled Pressure Regulation
- Multiple Programs for different parts, with the possibility of different pressures
- Multiple Languages, English, French, Spanish, Italian, …
- BarCode Reader inputs and Printer output
- Sequence – Pre-Fill to Balance Pressure, Reduced to Balanced Test Pressure, Stabilize, Measure Differential Pressure Decay
- Data Collection by Serial Number, using a PC or PLC ( optional Field buses, Ethernet IP, ProfiNet, EtherCAT Modbus)
Type | Pressure Decay |
# of Channels | 1 |
Resolution | 0.1 Pa (0.000015 psi) |
Display | 6.2 in Colour Display |
Data Storage | USB |
Interface | RS-232, RS-485, Profinet, EthernetIP, EtherCAT, Ethernet TCP/IP |
- Electronically Controlled Pressure Regulation
- Multiple Programs for different parts, with the possibility of different pressures
- Multiple Languages, English, French, Spanish, Italian, …
- BarCode Reader inputs and Printer output
- Sequence – Pre-Fill Pressure, Reduced to Test Pressure, Stabilize, Measure Absolute Pressure Decay
- Data Collection by Serial Number, using a PC or PLC ( optional Field buses, Ethernet IP, ProfiNet, EtherCAT Modbus)
- Dimensions 270 x 160 x 300 mm
- Weight 8 Kg
Type | Pressure Decay |
# of Channels | 2 |
Resolution | 0.1 Pa (0.000015 psi) |
Display | 6.5 in Color Display |
Data Storage | USB |
Interface | RS-232, RS-485, Profinet, EthernetIP, EtherCAT, Ethernet TCP/IP |
- Electronically Controlled Pressure Regulation
- Multiple Programs for different parts, with the possibility of different pressures
- Faster parameter programming
- Multiple Languages, English, French, Spanish, Italian, …
- Barcode Reader inputs and Printer output
- Sequence – Pre-Fill Pressure, Reduced to Test Pressure, Stabilize, Measure Absolute Pressure Decay
- Data Collection by Serial Number, using a PC or PLC ( optional Field buses, Ethernet IP, ProfiNet, EtherCAT Modbus )
Type | Pressure Decay |
# of Channels | 1 |
Resolution | 0.1 Pa (0.000015 psi) |
Display | 6.2 in Colour Display |
Data Storage | USB |
Interface | RS-232, RS-485, Profinet, EthernetIP, EtherCAT, Ethernet TCP/IP |