Loading...
A-Tech Instruments Ltd.
Français   |   Welcome. Please Register an Account or Login

4501, Pt100 Simulator

There are five precision decade switches in a sturdy metal housing. The desired temperature value is selected in four steps with a 0.1 degree celsius resolution in ranges from - 100 degree celsius to max. 500 degree celsius. According to DIN EN...
4501, Pt100 Simulator
4501, Pt100 Simulator
Features:
  • Simulation range -100 °C up to +500 °C
  • Resolution 0.1 °C
  • Calibration in accordance to DIN EN 60751
  • Simulation of line resistance 10 Ohms, 20 Ohms or 30 Ohms
  • Compact, easy to handle aluminium-housing
Manufacturer:  Burster Praezisionsmesstechnik Gmbh
Quote
Datasheet
Description
Specifications
Applications
Downloads
Similar Products
4501 - Description
There are five precision decade switches in a sturdy metal housing. The desired temperature value is selected in four steps with a 0.1 degree celsius resolution in ranges from - 100 degree celsius to max. 500 degree celsius. According to DIN EN 60751 the precision resistors simulate the temperature values for the Pt 100 resistor. The simulated temperature value is called on the output plugs "R". If required, the line resistance can also be simulated in steps of 10, 20 and 30 Ω.

The celsius scale, displaced by 273,15 K opposite the absolute temperature stipulates that an additional switch-over of polarity is performed at negative celsius temperature values. The simulator is high ohmic at wrongly entered + or signs. An unintentional misuse is practically impossible. The switches are implemented in a shortcircuit control manner. The precision resistors in the 100 deg. decade will therefore be switched parallelly at the moment of switch-over, in all other decade steps there is no effect at switchover.

The used resistor material MANGANIN® has a temperature coefficient smaller 10 ppm/K. This makes a consideration of the environmental temperature normally superfluous.
4501 - Specifications
Output TypeRTD
4501 - Applications
Wherever temperatures are measured, temperatures must also be simulated. The Pt-simulator is suitable for a wide area of applications. It has a wide range of simulation, which is resolved in 0.1 deg. C steps and makes many assignments in chemistry-, measuring-, controlling-, medicine and household applications, food industry, vehicle construction, air- and space travel and power plants easy to solve. Often in the past several simulators had to be used alongside to achieve either resolution or the range of the relevant application. As an extra advantage to the user temperatures can be entered in degrees celsius.

Additional extensive conversions and readings in tabulation sheets are no longer necessary
4501 - Downloads
4501 - Similar Products
  • 24 set temperature values
  • Calibration per DIN EN 60751
  • Resistor material: ZERANIN®
  • Temperature coefficient ≤ 2 ppm/K
  • Pt 100, Pt 500, Pt 1000
Output Type
RTD
  • Range: -200°C ...+850°C (Accuracy: 0.02°C)
  • Simulation of Pt100, Pt200,Pt500, Pt1000, Ni100 & Ni1000 Sensors
  • Temperature scales: ITS 90 & IPTS 68
  • Manual or remote control via RS232/IEEE488
  • A sturdy aluminium housing
Output Type
RTD
  • Wide Range of Inputs and Outputs all in one Digistant
  • Simulate and Measure
  • Vdc: 1mV up to 20V
  • Current: 1mA up to 24mA
  • Resistance: 5Ω up to 4000Ω & HART: 250Ω
  • Frequency/Pulse: 2 CPM up to 10KHz
  • RTD: Pt100 l 1000, Ni
  • TC: J, K, T, E, R, S, B, C
Input Type
DC Voltage,
DC Current,
Thermocouple,
RTD
Output Type
DC Voltage,
DC Current,
Thermocouple,
RTD
Interface
USB
Supply
Battery,
230 V AC
  • More than 60 fixed temperature values from - 200 C... + 850 C are in stock
  • Calibration in acc. with DIN EN 60751
  • +/- 0.02% tolerance for the models 1144-Pt 100, 1144-Pt 500 and 1144-Pt 1000
  • +/- 0.05% tolerance from the resistance value in the range - 200 C...- 50 C the model 1146-Pt 100
Output Type
RTD
  • DC voltage up to ±100.0000 V, ±0.002 %
  • DC current up to ±50.0000 mA, ±0.005 %
  • Thermocouple simulation of R, S, B, J, T, E, K, N, M, C, D, G2
  • 32 Automatic sequence functions ( ramp function, max. of 100 steps )
  • can be operated both via the keypad or via the Ethernet & USB interfaces
  • RTD Simulation Pt100 ... Pt1000, Ni100 ... Ni1000 (optional)
  • True ohmic resistance simulation 10 Ω ... 300 kΩ (optional)
  • Frequency simulation 0.01 Hz ... 15 kHz (optional)
Input Type
DC Voltage,
DC Current,
Thermocouple,
RTD
Output Type
DC Voltage,
DC Current,
Thermocouple,
RTD
Interface
USB,
Ethernet TCP/IP
Products    |    Applications    |    Services    |    Partners    |    About Us    |    Contact Us